15224
15228
CT BRW/Radiosurgery/CRW No IVcon IMG14023
Machine Model: Canon AQ 64
Generated:
2021-04-20 10:00:46
IV Contrast NA See IV Guidelines
for other access lines.
IV Saline NA
Oral Contrast NA
Water NA
Contrast amount determined by weight and renal function.
Special Instructions: IOML perpendicular to table (ring bolted to table), No Angle, 350 FOV, 1x1, Bottom ring through top ring
Scanogram: Lateral / AP
Landmark: Hyoid
A Non-Contrast
Scan Start Bottom BRW Ring
Scan Stop Top BRW Ring
Patient Position Supine
Scan Direction caudal/cranial
Scan Mode Helical
Scan Range If Volume NA
Scan Delay NA
Reconstruction 1 Thickness x Spacing 1 mm x 1 mm
Algorithm Brain
FOV 35 cm
FC 23
kVp 120 kVp
Scan FOV 35 cm
Pitch Detail
mA 240 mA
Rotation Time .75 sec
Detector Rows 0.5 mm x 64 mm
Indications: presurgical or radiosurgical planning for tumor, shunt placement, etc. , radiosurgery planning for vascular lesions .
0 images
0 images
0 images
0 images
Slide 0