| CTA Chest Pulmonary Vein with IV Contrast | IMG12130 | |
| Machine Model: | Canon AQ 64 |
Generated:
2021-02-10 06:45:34
|
|
|
|||||||||||||||
| Special Instructions: Above aortic arch through bottom of heart. | |||||||||
| Scanogram: AP/Lateral Landmark: Above Shoulders | |||||||||
| A | Non-Contrast | BT | Bolus Tracking | B | Arterial | C | Venous | ||
| Scan Start | Top of Heart | Center of PV | Above Aortic Arch | Aortic Arch | |||||
| Scan Stop | Bottom of Heart | Center of PV | Bottom of Heart | Above Aortic Arch | |||||
| Patient Position | Supine | Supine | Supine | Supine | |||||
| Scan Direction | Craniocaudal | NA | Craniocaudal | Craniocaudal | |||||
| Scan Mode | Helical | S&V | Helical | Helical | |||||
| Scan Range If Volume | NA | NA | NA | NA | |||||
| Scan Delay | NA | NA | 10 seconds | 10 seconds | |||||
| Bolus Tracking | NA | Pulm Vein 150 HU | Pulm Vein 150 HU | NA | |||||
| Reconstruction 1 | Thickness x Spacing | 1 mm x 1 mm | 1 mm x .8 mm | 1 mm x .8 mm | |||||
| Algorithm | Body STD | CTA Body | CTA Body | ||||||
| FOV | 40 cm | 40 cm | 40 cm | ||||||
| Reconstruction 2 | Thickness x Spacing | 1 mm x 1 mm | |||||||
| Algorithm | Lung STD | ||||||||
| FOV | RIB to RIB | ||||||||
| FC | 13 | 2 | 3 | 3 | |||||
| kVp | 120 kVp | 120 kVp | 120 kVp | 120 kVp | |||||
| Scan FOV | Patient Largest Width | Patient largest + 2CM | Patient largest + 2CM | Patient largest + 2CM | |||||
| Pitch | Standard | One Shot | Standard | Standard | |||||
| mA | modulated mA | 50 mA | modulated mA | modulated mA | |||||
| Rotation Time | .5 sec | .5 sec | .5 sec | .5 sec | |||||
| Detector Rows | 0.5 mm x 64 mm | 1 mm x 4 mm | 0.5 mm x 64 mm | 0.5 mm x 64 mm | |||||
| Indications: atrial fibrillation , pulmonary vein ablation , venous mapping prior to biventricular pacing . | |||||||||