| CT Hand No IVcon | IMG14012 | |
| Machine Model: | Canon Genesis 640 |
Generated:
2021-04-20 15:47:42
|
|
|
|||||||||||||||
| Special Instructions: Prone, Affected arm above head. | |||
| Scanogram: AP/Lateral Landmark: Area of Interest | |||
| A | Non-Contrast | ||
| Scan Start | Area of Interest | ||
| Scan Stop | Area of Interest | ||
| Patient Position | Prone | ||
| Scan Direction | Craniocaudal | ||
| Scan Mode | Helical | ||
| Scan Range If Volume | NA | ||
| Scan Delay | NA | ||
| Reconstruction 1 | Thickness x Spacing | 1 mm x 1 mm | |
| Algorithm | Body STD | ||
| FOV | 24 cm | ||
| Reconstruction 2 | Thickness x Spacing | .5 mm x .3 mm | |
| Algorithm | Bone | ||
| FOV | 24 cm | ||
| Reconstruction 3 | Thickness x Spacing | 2 mm x 1 mm | |
| Algorithm | Bone | ||
| FOV | Coronal + Sagittal | ||
| FC | 18 | ||
| kVp | Auto* kVp | ||
| Scan FOV | 24 cm | ||
| Pitch | Standard | ||
| mA | modulated mA | ||
| Rotation Time | .75 sec | ||
| Detector Rows | 0.5 mm x 80 mm | ||
| Indications: degenerative or posttraumatic etiology , trauma . | |||