| CT Ankle Joint with Hardware No IVcon | IMG12142 | |
| Machine Model: | Canon Genesis 640 |
Generated:
2021-04-19 21:24:09
|
|
|
|||||||||||||||
| Special Instructions: Supine, Affected ankle iso center, Unaffected leg/knee bent up out of FOV, Acquisition should extend at least 5 cm beyond full extent of hardware proximally and distally. Clinical indications may dictate greater coverage. | |||
| Scanogram: AP/Lateral Landmark: Area of Interest | |||
| A | Non-Contrast | ||
| Scan Start | Area of Interest | ||
| Scan Stop | Area of Interest | ||
| Patient Position | Supine | ||
| Scan Direction | Craniocaudal | ||
| Scan Mode | Helical | ||
| Scan Range If Volume | SEMAR | ||
| Scan Delay | NA | ||
| Reconstruction 1 | Thickness x Spacing | 1 mm x 1 mm | |
| Algorithm | Body STD | ||
| FOV | 32 cm | ||
| Reconstruction 2 | Thickness x Spacing | .5 mm x .3 mm | |
| Algorithm | Bone | ||
| FOV | 32 cm | ||
| Reconstruction 3 | Thickness x Spacing | .5 mm x .3 mm | |
| Algorithm | Bone SEMAR | ||
| FOV | 32 cm | ||
| Reconstruction 4 | Thickness x Spacing | 2 mm x 1 mm | |
| Algorithm | Bone | ||
| FOV | Coronal + Sagittal | ||
| FC | 18 | ||
| kVp | Auto* kVp | ||
| Scan FOV | 32 cm | ||
| Pitch | Standard | ||
| mA | modulated mA | ||
| Rotation Time | .75 sec | ||
| Detector Rows | 0.5 mm x 80 mm | ||
| Indications: degenerative or posttraumatic etiology , hardware assessment , trauma . | |||